The main use and working principle of VHP hydrogen peroxide transfer chamber
Time:2024-11-20      Click:13

The VHP sterilization transfer window is mainly used for the sterilization of the outer packaging and outer surface of the material to avoid the pollution caused by the material entering the high-level area from the low level area.

 

Configuration and performance

Low temperature environmental protection sterilization technology, hydrogen peroxide has good material compatibility, easy to decompose;

Fast sterilization, low operating cost;

Equipped with PLC and touch screen to control the operation of each stage, real-time monitoring of the effective operation of the equipment;

The independent ventilation and exhaust residue unit can quickly replace the hydrogen peroxide gas in the window to prevent it

Enter the HVAC system.

Internal with material support and track system, explosion-proof material handling.

 

Technical parameter

Model/Parameter

ATM288

ATM512

ATM1000

外尺寸(L x Wx H) mm

1200*800*2000

1400*800*2000

1600*1060*2100

内尺寸(Lx Wx H) mm

600x800x600

800x800x800

1000x1000x1000

Box material

Stainless steel plate

Give an alarm

High/low pressure alarm, open the door timeout, lock the door alarm

Residue level

<1ppm

Cleanliness of the cabin

A

Sterilizing agent

30% hydrogen peroxide solution

Biological decontamination cycle time

<90min

Biological decontamination effectiveness

6-10g kill rate

Filter grade

Air supply and exhaust efficiency H14EN1822)

Data integration function

Data saving and printing function

Power

3000w

3000w

3000w